Fluke Corporation MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 8026B: CAL ADJ /5100 DATE: 10-Feb-95 AUTHOR: Fluke Corporation REVISION: $Revision: 2.4 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 75 CONFIGURATION: Fluke 5100B ============================================================================= # # Source: # Fluke 8026B instruction manual (PN 646414 June 1982). # # Compatibility: # MET/CAL 3.0 or later # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5100B accuracy file contains 6 month specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 6 month specifications of the 5100B are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- J W 1.002 HEAD EQUIPMENT SETUP 1.003 DISP Ambient temperature: 21C - 25C. 1.003 DISP Relative humidity: less than 80%. 1.003 DISP Allow the UUT 30 min. to stabilize. 1.004 HEAD CALIBRATION ACCESS PROCEDURE 1.005 DISP Disconnect the test leads and the battery eliminator 1.005 DISP if attached. Open the battery compartment, and 1.005 DISP disconnect the battery. Remove the three phillips-head 1.005 DISP screws from the bottom of the case. Turn the instrument 1.005 DISP face-up, and grasp the top cover at both sides of the 1.005 DISP input connectors. Then, pull the top cover from the UUT. 1.005 DISP Re-connect the battery and turn the UUT on. 1.006 HEAD DC VOLTAGE CALIBRATION ADJUSTMENT 1.007 DISP Connect the 5100B and UUT as follows: 1.007 DISP [32] 5100B OUTPUT HI to UUT V Ohm S 1.007 DISP [32] 5100B OUTPUT LO to UUT COMMON 1.008 RNG 200mV 1.009 5100 190mV S 2W 1.010 DISP Adjust R5 (DC CAL) for a display of 190.0. 1.011 5100 * S 1.012 HEAD AC VOLTAGE CALIBRATION ADJUSTMENT 1.013 RNG 200mV AC 1.014 5100 190mV 100H S 2W 1.015 DISP Adjust the AC CAL potentiometer marked "F" (fine 1.015 DISP adjustment- R19) so that it is at mid-range. 1.015 DISP 1.015 DISP Adjust the AC CAl potentiometer marked "C" 1.015 DISP (coarse-adjust-R17) for a UUT display of 1.015 DISP 190.0 +/-3 counts 1.016 DISP Adjust R19 ("F" fine adjust) for a UUT display of 1.016 DISP 190.0 or 190.1. 1.017 5100 * S 1.018 RNG 20V AC 1.019 5100 19V 7kH S 2W 1.020 DISP Adjust C9 for a display of 19.00V (+/-2 digits). 1.021 5100 * S 1.022 RNG 2V AC 1.023 5100 1.9V 10kH S 2W 1.024 DISP Adjust C2 for a display of 1.900V (+/-2 digits). 1.025 END